Common Questions and Answers about Tensile Testing Machines
1.The power of the main unit of the testing machine is not on, and the equipment cannot move up and down:
A) Check whether the power line connected to the testing machine is properly connected.
B) Check whether the emergency stop switch is in the open position;
C) Check whether the power voltage connected to the testing machine is normal;
D) Check if the fuse on the machine socket has blown. If so, remove the original fuse and install a spare one instead.
2. The main unit is powered on, but the testing machine does not operate:
A) If the power supply voltage is abnormal, please install an UPS power supply.
B) The communication line between the computer and the main unit is not properly connected. Tighten and secure the plug.
C) The communication serial port of the computer is damaged. A new communication serial port has been replaced.
D) When the limit switch operates, adjust the limit stop ring to the appropriate position.
3. If the sample is not completely torn, peeled off or stretched apart, the test will be stopped.
A) The setting of the judgment conditions in the "test plan" of the software is incorrect. It can be corrected by making the necessary changes.
B) When the force value of the test reaches the maximum force of the testing machine, the sample should be sawn off to prevent damage to the sensor.
4. No data was provided when the report was prepared:
Confirm whether all the "parameters" have been entered correctly. Additionally, click "Save Parameters" (such as diameter, width, thickness, pitch, etc.); if no saving is performed after clicking "Results", and whether the corresponding report template has been selected